Wet column generator elution and differences from the dry column
Awaiting physician review
To obtain sterile pertechnetate from a Mo-99/Tc-99m generator with a permanently connected eluent reservoir (wet column) and to perform correctly the steps that differ from a dry column.
What you need
- The product's evacuated elution vials (the vacuum volume sets the elution volume)
- Sterile needle caps supplied with the product
- Elution vial shield with lead cap
- Disinfectant recommended by the manufacturer
- Gloves, lab coat, body and ring dosimeters
- Dose calibrator, Mo-99 and Al³⁺ test materials, record form, label
Steps
- Put on your dosimeters and gloves. Check the generator's calibration and expiry dates (20 days from manufacture for one product) and the time of the last elution.
- Check that the eluent reservoir (saline bag) is connected to the column; in a wet system no separate saline vial is fitted for elution.
- Choose the evacuated vial for the volume you need (a 5 mL vial for one product; 10 or 15 mL can also be used). Place it in its shield and wipe the stopper with the disinfectant the manufacturer recommends.
- Remove the sterile cap from the elution needle and push the shielded vial fully onto the needle. The vacuum draws saline from the reservoir through the column into the vial; do not move the vial until the flow stops.
- Remove the vial and fit its lead cap. Immediately place a new sterile needle cap on the elution needle: the needle is never left open between elutions.
- Assay the eluate in the dose calibrator on the Tc-99m setting; record the time, volume and activity and calculate the radioactive concentration.
- Quality control: appearance, Mo-99 breakthrough and Al³⁺ test (separate card). If acceptable, label the vial.
- Store and use the eluate as the product information specifies (for one product at 2-8 °C and within 10 hours of elution).
Key points
- In a wet system a large saline reservoir stays connected to the inlet and the column remains full of saline after elution; in a dry system a separate saline vial is fitted for each elution and the column is emptied. The volume is set by the evacuated vial in a wet system and by the saline vial in a dry system.
- Radiolysis of the water left on the column produces reducing species; reduced technetium binds more strongly to the alumina and the elution yield falls. This is why dry systems are more widely used.
- With calibrated vials for 3 mL fractionated elution, the volume was more variable and the yield slightly lower with a wet column (median 81.8%, vs 84.7% with a dry column).
- For short-interval elutions use the product's table of obtainable activity: for one product 39% at 4 hours and 62% at 8 hours. These are lower than theoretical ingrowth (46% at 4 hours, 73% at 8 hours).
- For one product the first eluate can be used normally; for kits that need fresh eluate the kit's product information applies. For the same product, the stopper and needle must not be wiped with ethanol or ether, as this may disturb elution. The current product information of the generator in use applies.
- The example product on this card (ELUMATIC III) is no longer in the manufacturer's current product list; the product-specific values come from the last available product information. The same manufacturer's current Tc-99m generators are dry-column (a saline vial is fitted for each elution); if you meet a wet-column generator, use that generator's own product information.
Common mistakes
- Using a disinfectant the product prohibits (e.g. ethanol)
- Leaving the needle uncapped after elution
- Eluting an unwanted volume with a vial of the wrong vacuum volume
- Removing the vial before the flow has stopped
- Looking for or trying to fit a saline vial out of dry-column habit
Safety. Use the shielded vial and tongs; keep hands-on time short; the column and needle are never removed from their shielding.
Sources: EANM 2019 (Technologist's Guide, Radiopharmacy: an update, Bölüm 4 jeneratörler) · SmPC: ELUMATIC III (CIS bio; son erişilebilir metin İngiltere 09.11.2016 ve İrlanda 2017; İrlanda ruhsatı 31.12.2017'de geri çekildi, MHRA ve Fransa ANSM kayıtlarında ve Curium ürün listesinde 2026'da yok) · Curium ürün listesi 2026 (güncel Tc-99m jeneratörleri kuru kolon: Ultra-Technekow FM, Türkiye SmPC 21.08.2024; MONTEK, Almanya Fachinformation 09.2025) · Vigne ve ark., Nucl Med Commun 2018;39:865-869 (doi 10.1097/MNM.0000000000000881, PMID 29975214). Last reviewed: 23 November 2026.