  1. 10.1 Dosimetric Quantities
  2. 10.1.1 Absorbed Dose
  3. 10.1.2 Equivalent Dose
  4. 10.1.3 Effective Dose
  5. 10.1.4 Exposure
  6. 10.1.5 Kerma (Kinetic Energy Released per unit Mass)
  7. 10.1.7 Operational Dose Quantities
  8. 10.1.6 Summary Table of Dosimetric Quantities
  9. 10.2 The MIRD Formalism
  10. 10.2.1 The Basic MIRD Equation
  11. 10.2.2 Cumulated Activity (Ã – A tilde)
  12. 10.2.3 S Value (S Factor)
  13. 10.2.4 Residence Time (τ)
  14. 10.2.5 Steps of Dose Calculation with the MIRD Formalism
  15. 10.2.6 OLINDA/EXM Software
  16. 10.3 Dose Limits (ICRP Recommendations)
  17. 10.3.1 Basic ICRP Principles
  18. 10.3.2 Occupational Dose Limits (ICRP 103)
  19. 10.3.3 Public Dose Limits (ICRP 103)
  20. 10.3.4 Dose Limits for the Pregnant Radiation Worker
  21. 10.3.5 Embryo/Fetus Dose Limits and Threshold Values
  22. 10.4 Personal Dosimetry and Monitoring
  23. 10.4.1 Types of Personal Dosimeter
  24. 10.4.2 Rules for Wearing Dosimeters
  25. 10.4.3 Biological Dosimetry
  26. 10.5 Radiation Safety Practice
  27. 10.5.1 The Three Basic Principles of Radiation Protection
  28. 10.5.2 Components of a Radiation Safety Programme
  29. 10.5.3 Security of Radioactive Material
  30. 10.5.4 Area Classification
  31. 10.5.5 Specific Gamma-Ray Constant (Γ – Exposure Rate Constant)
  32. 10.6 Contamination and Decontamination
  33. 10.6.1 Types of Contamination
  34. 10.6.2 Detection of Contamination
  35. 10.6.3 Decontamination Procedures
  36. 10.6.4 Internal Contamination and Its Treatment
  37. 10.7 Radiation Safety in Pregnancy and Breastfeeding
  38. 10.7.1 Diagnostic Nuclear Medicine in Pregnancy
  39. 10.7.2 Radioiodine Exposure in Pregnancy
  40. 10.7.3 Radiation Safety in Breastfeeding
  41. 10.7.4 Pregnancy Screening Before a Nuclear Medicine Study
  42. 10.8 Patient Release Criteria (After RAI Therapy)
  43. 10.8.1 NRC Release Criteria (10 CFR 35.75)
  44. 10.8.2 Patient Instructions After Release
  45. 10.8.3 Death of a Radioactive Patient
  46. 10.8.4 Reporting of Misadministration
  47. Key Points for the Board
  48. Current guidelines on this chapter's topics
← Back‹ Chapter 9Chapter 11 ›

Chapter 10Basic sciencesLast reviewed: 14 Nov 2026 · What's new

Radiation Dosimetry and Safety

This chapter's English text is a draft translation (prepared with Claude and checked against a terminology glossary and automatic checks); specialist review is in progress. Please report errors with the feedback button.

Solve questions (24) →
Contents
  1. 10.1 Dosimetric Quantities
  2. 10.1.1 Absorbed Dose
  3. 10.1.2 Equivalent Dose
  4. 10.1.3 Effective Dose
  5. 10.1.4 Exposure
  6. 10.1.5 Kerma (Kinetic Energy Released per unit Mass)
  7. 10.1.7 Operational Dose Quantities
  8. 10.1.6 Summary Table of Dosimetric Quantities
  9. 10.2 The MIRD Formalism
  10. 10.2.1 The Basic MIRD Equation
  11. 10.2.2 Cumulated Activity (Ã – A tilde)
  12. 10.2.3 S Value (S Factor)
  13. 10.2.4 Residence Time (τ)
  14. 10.2.5 Steps of Dose Calculation with the MIRD Formalism
  15. 10.2.6 OLINDA/EXM Software
  16. 10.3 Dose Limits (ICRP Recommendations)
  17. 10.3.1 Basic ICRP Principles
  18. 10.3.2 Occupational Dose Limits (ICRP 103)
  19. 10.3.3 Public Dose Limits (ICRP 103)
  20. 10.3.4 Dose Limits for the Pregnant Radiation Worker
  21. 10.3.5 Embryo/Fetus Dose Limits and Threshold Values
  22. 10.4 Personal Dosimetry and Monitoring
  23. 10.4.1 Types of Personal Dosimeter
  24. 10.4.2 Rules for Wearing Dosimeters
  25. 10.4.3 Biological Dosimetry
  26. 10.5 Radiation Safety Practice
  27. 10.5.1 The Three Basic Principles of Radiation Protection
  28. 10.5.2 Components of a Radiation Safety Programme
  29. 10.5.3 Security of Radioactive Material
  30. 10.5.4 Area Classification
  31. 10.5.5 Specific Gamma-Ray Constant (Γ – Exposure Rate Constant)
  32. 10.6 Contamination and Decontamination
  33. 10.6.1 Types of Contamination
  34. 10.6.2 Detection of Contamination
  35. 10.6.3 Decontamination Procedures
  36. 10.6.4 Internal Contamination and Its Treatment
  37. 10.7 Radiation Safety in Pregnancy and Breastfeeding
  38. 10.7.1 Diagnostic Nuclear Medicine in Pregnancy
  39. 10.7.2 Radioiodine Exposure in Pregnancy
  40. 10.7.3 Radiation Safety in Breastfeeding
  41. 10.7.4 Pregnancy Screening Before a Nuclear Medicine Study
  42. 10.8 Patient Release Criteria (After RAI Therapy)
  43. 10.8.1 NRC Release Criteria (10 CFR 35.75)
  44. 10.8.2 Patient Instructions After Release
  45. 10.8.3 Death of a Radioactive Patient
  46. 10.8.4 Reporting of Misadministration
  47. Key Points for the Board
  48. Current guidelines on this chapter's topics
Learning objectives and modules3 modules · about 2 h 25 min

Learning objectives

After this chapter you will be able to:

  • Distinguish absorbed dose, equivalent dose, effective dose and the operational quantities with their units.
  • Calculate absorbed dose with the MIRD formalism: cumulated activity, effective half-life and S value.
  • State occupational and public dose limits, worker categories and area classification.
  • Explain personal dosimetry methods and correct use of the dosimeter.
  • List the steps to follow after a spill or contamination.
  • Apply the criteria for radiopharmaceutical administration in pregnancy and breastfeeding and for patient release after radioiodine therapy.

Modules

The chapter is divided into study modules without changing the order of the text: read, watch the animations, then answer the questions linked to those headings.

  1. Module 110.1–10.3 Dosimetric Quantities · The MIRD Formalism · Dose Limits (ICRP Recommendations)55 min · 1 animation · 22 questionsAnswer the questions →
  2. Module 210.4–10.5 Personal Dosimetry and Monitoring · Radiation Safety Practice45 min · 1 animation · 19 questionsAnswer the questions →
  3. Module 310.6–10.8 Contamination and Decontamination · Radiation Safety in Pregnancy and Breastfeeding · Patient Release Criteria (After RAI Therapy) · Key Points for the Board45 min · 15 questionsAnswer the questions →

Times are estimates: 130 words per minute, 3 min per animation, 1.5 min per question.

Tip cards (42) →

MOST FREQUENTLY ASKED IN THIS CHAPTER

✓
Effective dose cannot be measured → H*(10) (ambient), Hp(10) (whole body), Hp(0.07) (skin), Hp(3) (eye lens)
✓
Controlled area vs supervised area
✓
Category A: likely to receive >6 mSv/year effective, >15 mSv/year eye lens, >150 mSv/year skin/extremity; Category B: those who may exceed the public limit
✓
Absorbed dose: Gray (Gy) = 100 rad → a physical quantity, does not take radiation type into account
✓
Equivalent dose: Sievert (Sv) = 100 rem → corrected with the radiation weighting factor (wR)
✓
Effective dose: Sievert (Sv) → corrected with the tissue weighting factor (wT) → indicator of stochastic risk
✓
For photons and beta wR = 1 → absorbed dose (Gy) = equivalent dose (Sv) numerically

10.1 Dosimetric Quantities

Radiation dosimetry covers the quantities used to measure and evaluate the effect of radiation on biological tissue. In nuclear medicine, a thorough understanding of these concepts is critical for both patient dosimetry and staff dosimetry.

10.1.1 Absorbed Dose

Definition: The amount of energy imparted to tissue by radiation per unit mass.

Formula:

D = dE / dm

•
D: Absorbed dose
•
dE: Energy imparted (joule)
•
dm: Mass (kg)

Unit:

•
SI unit: Gray (Gy) → 1 Gy = 1 J/kg
•
Old unit: rad → 1 rad = 0.01 Gy = 1 cGy
•
Conversion: 1 Gy = 100 rad

Properties:

•
A purely physical quantity → does not take radiation type or biological effect into account
•
Applies to all types of radiation (alpha, beta, gamma, neutron)
•
Independent of tissue type
•
The basic building block of dosimetric calculations
BOARD TIP
▸
"1 Gy = ? rad"

→ 100 rad.

▸
"1 rad = ? Gy"

→ 0.01 Gy = 1 cGy. These conversions are asked frequently in the exam. Remember the 100-fold difference between Gy and rad!

10.1.2 Equivalent Dose2 questions

Definition: The absorbed dose corrected for the biological effectiveness of the radiation type. It reflects the fact that different radiation types cause different biological damage at the same absorbed dose.

Formula:

HT = ΣR wR × DT,R

•
HT: Equivalent dose in organ T
•
wR: Radiation weighting factor
•
DT,R: Absorbed dose to organ T from radiation type R

Unit:

•
SI unit: Sievert (Sv) → 1 Sv = 1 J/kg (biologically weighted)
•
Old unit: rem → 1 rem = 0.01 Sv
•
Conversion: 1 Sv = 100 rem
•
wR values are derived from RBE data for stochastic effects at low doses; the quality factor Q (formerly QF) used in operational quantities is defined as a function of LET

Radiation Weighting Factors (wR) – ICRP 103:

Radiation TypewR
Photons (X-rays, gamma)1
Electrons (beta)1
Positrons1
Protons2
Alpha particles20
NeutronsEnergy-dependent (2.5–20)
Heavy ions20
BOARD TIP
▸
Memorise the radiation weighting factors: Photon/beta/positron = 1, Proton = 2, Alpha = 20, Neutron = energy-dependent (2.5–20). In the exam, "wR of an alpha particle?" → 20.
▸
"wR of gamma rays?"

→ 1. Because wR = 1 for photons and beta, for these radiations the absorbed dose (Gy) = equivalent dose (Sv) numerically.

10.1.3 Effective Dose4 questions

Definition: The total body dose calculated by taking into account the radiation sensitivity of the different organs and tissues of the body. It is a general indicator of stochastic risk (cancer, hereditary effects).

Formula:

E = ΣT wT × HT

•
E: Effective dose
•
wT: Tissue weighting factor
•
HT: Equivalent dose in organ T

Unit: Sievert (Sv) – the same unit as equivalent dose but with a different meaning

Tissue Weighting Factors (wT) – ICRP 103:

Tissue/OrganwT
Bone marrow (red)0.12
Colon0.12
Lung0.12
Stomach0.12
Breast0.12
Gonads0.08
Bladder0.04
Oesophagus0.04
Liver0.04
Thyroid0.04
Bone surface0.01
Brain0.01
Salivary glands0.01
Skin0.01
Remainder tissues0.12
TOTAL1.00

Key points:

•
The sum of the wT values must always be 1.00
•
With ICRP 103 (2007), the gonad wT value was reduced from 0.20 to 0.08 (compared with ICRP 60)
•
The breast wT value was raised from 0.05 to 0.12
•
Tissues with the highest wT value (0.12 each): bone marrow, colon, lung, stomach, breast, remainder tissues
•
Collective effective dose: The sum of the effective doses of individuals in a population (unit person·Sv / man·Sv). It is used to compare the burden of medical exposure on society; it should not be used to estimate individual risk
•
Effective dose per unit activity in adults: Tl-201 ~0.14-0.22, Ga-67 ~0.10, In-111 DTPA ~0.02, most Tc-99m agents ~0.005-0.01 mSv/MBq. Single-photon agents with long half-lives give a much higher dose per MBq
BOARD TIP
▸
"Organs with the highest wT value in ICRP 103?"

→ Bone marrow, colon, lung, stomach, breast (0.12 each).

▸
"How did the gonad wT value change from ICRP 60 to 103?"

→ 0.20 → 0.08 (decreased).

▸
"Breast wT value?"

→ 0.12 (in ICRP 60 it was 0.05 → increased). These changes are asked frequently in the exam!

10.1.4 Exposure

Definition: The amount of ionisation produced in air by X-rays or gamma rays. It is defined only for photons and measured only in air.

Unit:

•
SI unit: C/kg (coulomb/kilogram)
•
Old unit: Roentgen (R) → 1 R = 2.58 × 10⁻⁴ C/kg
•
In air 1 R ≈ 0.87 rad; in tissue, absorbed dose = f × exposure. The f factor (rad/R) is ~0.9-0.97 in soft tissue, whereas in bone it rises markedly at low photon energies because of the Z³ dependence of the photoelectric effect

Limitations:

•
Valid only for photons (X and gamma) → not used for particles
•
Defined only in air → not a tissue dose
•
Defined only for photons below 3 MeV
•
Its use has declined in modern dosimetry → absorbed dose and effective dose are preferred

10.1.5 Kerma (Kinetic Energy Released per unit Mass)

Definition: The kinetic energy of charged particles released per unit mass.

Formula:

K = dEtr / dm

Unit: Gray (Gy)

Relationship to absorbed dose:

•
Under conditions of electronic equilibrium: Kerma ≈ Absorbed dose
•
Without electronic equilibrium (e.g. at interfaces): Kerma ≠ Absorbed dose
•
Kerma measures the energy transferred by photons to the medium; absorbed dose measures the energy deposited in the medium

10.1.7 Operational Dose Quantities3 questions

Effective dose cannot be measured directly, even in a uniform external field. For this reason, "operational" quantities are used to calibrate radiation protection instruments:

•
Ambient dose equivalent H*(10): The equivalent dose at a depth of 10 mm in a tissue-equivalent material (unit Sv); area monitors and survey meters work with this quantity
•
Personal dose equivalent Hp(d): The equivalent dose in soft tissue at depth d below a specified point on the body (unit Sv); personal dosimeters work with this quantity
•
Hp(10) is used to estimate effective dose, Hp(0.07) to assess skin (and extremity) dose, and Hp(3) (d = 0.3 cm) to assess eye lens dose
BOARD TIP
▸
Effective dose is calculated, not measured. Measuring instruments are calibrated in operational quantities: H*(10) → ambient/area; Hp(10) → whole body (estimate of effective dose); Hp(0.07) → skin; Hp(3) → eye lens.
personal dosimeter body surface depth d = 0.07 mm Hp(0.07) skin and extremity dose d = 3 mm Hp(3) eye lens dose d = 10 mm Hp(10) whole body → effective dose estimate Personal dose equivalent tissue-equivalent medium (not to scale) H*(10): ambient dose equivalent → area monitor, survey meter

Figure 10.1: Measurement depths of the operational dose quantities: Hp(0.07) skin, Hp(3) eye lens, Hp(10) whole body (schematic).

Did this figure help you understand the topic?

10.1.6 Summary Table of Dosimetric Quantities

QuantitySI UnitOld UnitConversionWhat It Measures
Absorbed DoseGray (Gy)rad1 Gy = 100 radPhysical energy transfer
Equivalent DoseSievert (Sv)rem1 Sv = 100 remBiological effectiveness (radiation type)
Effective DoseSievert (Sv)rem1 Sv = 100 remStochastic risk (organ sensitivity)
ExposureC/kgRoentgen (R)1 R = 2.58×10⁻⁴ C/kgIonisation in air (photons only)
KermaGray (Gy)rad1 Gy = 100 radKinetic energy released
BOARD TIP
▸
To remember the unit conversions: the "Gy-rad" and "Sv-rem" pairs → both differ by a factor of 100 (1 Gy = 100 rad, 1 Sv = 100 rem). The exposure unit roentgen is valid only for photons and in air → not used for particles or tissue!
Absorbed dose (D) D = dE / dm Gy (old: rad) physical energy transfer (independent of radiation type) Equivalent dose (HT) HT = Σ wR × DT,R Sv (old: rem) biological effectiveness (by radiation type) Effective dose (E) E = Σ wT × HT Sv (old: rem) stochastic risk (by organ sensitivity) × wR × wT, Σ Radiation weighting factor wR photon (X, γ) 1 electron (β), positron 1 proton 2 alpha, heavy ions 20 neutron 2.5–20 (energy-dependent) Tissue weighting factor wT (ICRP 103; total 1.00) 0.12 bone marrow, colon, lung, stomach, breast, remainder tissues 0.08 gonads (ICRP 60: 0.20) 0.04 bladder, oesophagus, liver, thyroid 0.01 bone surface, brain, salivary glands, skin For photons and β wR = 1 → Gy and Sv numerically equal

Figure 10.2: From absorbed dose to equivalent and effective dose: radiation (wR) and tissue (wT) weighting factors.

Did this figure help you understand the topic?

10.2 The MIRD Formalism

The MIRD (Medical Internal Radiation Dose) formalism is the standard method used to calculate the internal radiation dose from radiopharmaceuticals administered to the body. It forms the basis of patient dosimetry in nuclear medicine.

10.2.1 The Basic MIRD Equation1 questions

General formula:

D(rT) = ΣrS Ã(rS) × S(rT ← rS)

•
D(rT): Absorbed dose in the target organ (Gy)
•
Ã(rS): Cumulated activity in the source organ (Bq·s or µCi·h)
•
S(rT ← rS): S value – dose to the target organ per unit cumulated activity in the source organ (Gy/Bq·s)

10.2.2 Cumulated Activity (Ã – A tilde)4 questions

Definition: Represents the total number of radioactive decays in an organ. It is the area under the activity–time curve.

Formula (mono-exponential model):

à = A₀ × f × Teff / ln 2 = 1.443 × A₀ × f × Teff

•
A₀: Total injected activity
•
f: Fraction of activity going to the organ (uptake fraction)
•
Teff: Effective half-life

Effective half-life:

1/Teff = 1/Tphys + 1/Tbiol or Teff = (Tphys × Tbiol) / (Tphys + Tbiol)

•
Tphys: Physical half-life (radioactive decay)
•
Tbiol: Biological half-life (biological elimination)
•
Teff is always shorter than both Tphys and Tbiol
BOARD TIP
▸
"Is the effective half-life closer to the physical or the biological half-life?"

→ Closer to the shorter one. Example: Tphys = 6 hours, Tbiol = 24 hours → Teff = (6×24)/(6+24) = 144/30 = 4.8 hours (closer to the physical half-life). Teff calculations are asked frequently in the exam!

Atoms in an organ are lost by physical decay or biological excretion; physical, biological and effective retention curves and the area under the activity–time curve (cumulated activity) are shown.

Continuous stream · no seeking

Key point: the effective half-life is shorter than both the physical and biological half-lives (1/Teff = 1/Tphys + 1/Tbiol); cumulated activity is the area under the curve, Ã = 1.443 × A₀ × Teff.

Source: book (example Tphys 6 h, Tbiol 24 h → Teff 4.8 h); A₀ = 100 MBq and f = 1 schematic; Tc-99m 6.01 h, I-123 13.2 h, I-131 8.02 d.

How to use: one click on the organ scene pauses and resumes; move over or drag along the graph; in the free step ↑ ↓ change the biological half-life smoothly; ← → change steps.

Speed
Test yourself:Chapter 10 · question 4Chapter 10 · question 5
Did this animation help you understand the topic?

Figure 10.3: Cumulated activity as the area under the curve (A) and the effective half-life being shorter than the physical and biological half-lives (B).

10.2.3 S Value (S Factor)1 questions

Definition: The absorbed dose delivered to the target organ per unit cumulated activity in the source organ.

Factors determining the S value:

•
Emission characteristics of the radionuclide (energy, type, frequency)
•
Geometric relationship between source and target organs (distance, size)
•
Masses of the source and target organs
•
Attenuation properties of the intervening tissues
•
S = Σ Δᵢ·φᵢ / m(target): Δ denotes the mean emitted energy and φ the absorbed fraction; φ depends on photon energy, the size and shape of the organ, and the distance between source and target. For the self-dose of non-penetrating β radiation φ ≈ 1

S value tables:

•
Calculated and published by the MIRD committee for standard phantoms (male, female, paediatric)
•
Separate S values are available for different radionuclides and source–target organ pairs
•
Can be calculated with OLINDA/EXM software

Key points:

•
Self-dose: When source and target organ are the same → S(rT ← rT)
•
Cross-dose: When source and target organ are different → S(rT ← rS)
•
Beta particles have a short range → self-dose dominates, cross-dose is negligible
•
Gamma rays have a long range → cross-dose is important

10.2.4 Residence Time (τ)

Definition: The ratio of cumulated activity to injected activity.

Formula:

τ = Ã / A₀

•
τ: Residence time (hours or seconds)
•
Ã: Cumulated activity
•
A₀: Injected activity

Clinical importance:

•
Residence time reflects how long the radiopharmaceutical stays in the organ
•
Long residence time → high organ dose
•
In dosimetric calculations τ can be used instead of Ã: D = A₀ × τ × S

10.2.5 Steps of Dose Calculation with the MIRD Formalism

1. Determine the biodistribution of the radiopharmaceutical: How much activity goes to which organs (uptake fraction)

2. Construct the activity–time curve for each organ: With serial imaging or blood/urine samples 3. Calculate cumulated activity (Ã): Area under the curve (integral)

4. Select the appropriate S values: From MIRD tables for the radionuclide and organ pair 5. Calculate the dose for each target organ: D = Σ Ã × S (sum of contributions from all source organs)

6. Calculate the effective dose: E = Σ wT × HT

10.2.6 OLINDA/EXM Software

•
Internal dosimetry calculation software based on the MIRD formalism
•
Includes standard phantoms (adult male/female, paediatric age groups)
•
S values are available for more than 800 radionuclides
•
Organ doses and effective dose are calculated by entering residence times
•
The standard tool in the dosimetric evaluation of new FDA-approved radiopharmaceuticals
BOARD TIP
▸
Remember the MIRD formula simply: Dose = Cumulated activity × S value. Cumulated activity answers "how much radioactivity stayed in the organ, and for how long". The S value answers "how much dose per unit activity". Their product gives the total dose.
Source organ (rS) Target organ (rT) γ: long range → cross-dose important S(rT ← rS): cross-dose Ã(rS): cumulated activity β: short range → self-dose dominates S(rS ← rS): self-dose D(rT): absorbed dose MIRD equation D(rT) = Σ Ã(rS) × S(rT ← rS) τ = Ã / A0 D = A0 × τ × S Biodistribution (uptake fraction) 1 Activity–time curves 2 Ã = area under the curve 3 S values (MIRD tables) 4 D = Σ Ã × S (each target organ) 5 E = Σ wT × HT 6

Figure 10.4: The MIRD formalism: self-dose and cross-dose from source to target organ and the steps of dose calculation (schematic).

Did this figure help you understand the topic?

10.3 Dose Limits (ICRP Recommendations)

In radiation protection, dose limits are set to keep the stochastic and deterministic risks of individuals exposed occupationally or as members of the public at an acceptable level.

10.3.1 Basic ICRP Principles1 questions

The ICRP (International Commission on Radiological Protection) defines three basic principles of radiation protection:

1. Justification:

–
A radiation practice should be carried out only if it provides a net benefit
–
The benefit must outweigh the risk
–
Each practice should be assessed individually

2. Optimisation (ALARA):

–
Doses should be kept as low as reasonably achievable
–
ALARA: As Low As Reasonably Achievable
–
Optimisation takes economic and social factors into account

3. Dose Limitation:

–
Individual doses should not exceed the set limits
–
Dose limits are defined separately for occupational and public exposure
–
Dose limits do not apply to medical exposure (patient dose)
–
Comparison: The world average from natural radiation is ~2.4 mSv/year (UNSCEAR). The largest share is radon (~1.3), followed by terrestrial (~0.5), cosmic (~0.4) and internal (~0.3 mSv) sources. Natural and medical exposures are not included in the dose limits
BOARD TIP
▸
"To whom do dose limits apply?"

→ Radiation workers and the public.

▸
"Do dose limits apply to patients?"

→ NO → patient doses are managed by the principles of justification and optimisation; there is no fixed dose limit. This distinction is asked frequently in the exam!

10.3.2 Occupational Dose Limits (ICRP 103)5 questions

Dose TypeLimitPeriod
Effective dose50 mSv (single-year maximum)Single year
Effective dose100 mSv / 5 years (average 20 mSv/year)5-year period
Equivalent dose to the eye lens20 mSv/year (average, 5 years)Annual average
Equivalent dose to the eye lens50 mSv/year (single-year maximum)Single year
Equivalent dose to the skin500 mSv/yearAnnual
Equivalent dose to the hands and feet500 mSv/yearAnnual

Important notes:

•
The eye lens limit was reduced by ICRP 118 (2012) from 150 mSv/year to 20 mSv/year → because of cataract risk
•
Over a 5-year period the total effective dose should not exceed 100 mSv
•
It should not exceed 50 mSv in any single year

10.3.3 Public Dose Limits (ICRP 103)1 questions

Dose TypeLimitPeriod
Effective dose1 mSv/yearAnnual
Equivalent dose to the eye lens15 mSv/yearAnnual
Equivalent dose to the skin50 mSv/yearAnnual

10.3.4 Dose Limits for the Pregnant Radiation Worker

•
Fetal dose during pregnancy (for the remainder of the pregnancy): ≤1 mSv (ICRP recommendation)
•
NRC (USA): Fetal dose during pregnancy ≤5 mSv (500 mrem)
•
Monthly fetal dose after declaration of pregnancy: ≤0.5 mSv (NCRP recommendation)
•
A pregnant worker may continue to work with radiation, but the dose limits must be observed
•
Declaration of pregnancy is voluntary (not mandatory)
BOARD TIP
▸
Memorise the dose limits side by side: Occupational = 20 mSv/year (5-year average, single year ≤50), Public = 1 mSv/year, Fetal = 1 mSv (ICRP) or 5 mSv (NRC) for the whole pregnancy. The occupational limit is 20 times the public limit. The eye lens limit was reduced by ICRP 118 from 150 → 20 mSv/year → this update may be asked in the exam!
1 10 100 1000 Dose limit (mSv/year, logarithmic scale) Effective dose 20 mSv (5-yr avg.; single year ≤50) 1 mSv Eye lens 20 mSv (5-yr avg.; single year ≤50) 15 mSv Skin 500 mSv 50 mSv Hands and feet 500 mSv no separate public limit occupational (ICRP 103) public (ICRP 103) Occupational effective dose limit is 20 times the public limit Fetus of a pregnant worker: ≤1 mSv (ICRP, rest of pregnancy) / ≤5 mSv (NRC, whole pregnancy) Occupational eye lens limit 150 → 20 mSv/year with ICRP 118.

Figure 10.5: Comparison of ICRP 103 occupational and public dose limits.

Did this figure help you understand the topic?

10.3.5 Embryo/Fetus Dose Limits and Threshold Values

EffectThreshold DoseGestational Period
Lethal effect (death)~100 mGyPreimplantation (0-2 weeks)
Major congenital anomaly~100-200 mGyOrganogenesis (2-8 weeks)
Mental retardation (severe)~300 mGy8-15 weeks (most sensitive)
Mental retardation (severe)≥300 mGy, probably higher (~500-700 mGy); risk ~4 times lower than at 8-15 weeks16-25 weeks (less sensitive)
IQ loss~25 IQ points/Gy8-15 weeks
Microcephaly~200 mGy8-15 weeks
Childhood cancerNo threshold (stochastic)All periods

(For details see Section 4.7 – Radiation Effects in Pregnancy)

Organogenesis Early fetal (most sensitive) Fetal (less sensitive) Late fetal Preimplantation (0–2 wk): lethal effect ~100 mGy Major anomaly ~100–200 mGy Severe mental retardation ~300 mGy · IQ loss ~25 points/Gy · microcephaly ~200 mGy Severe mental retardation ≥300 mGy (probably ~500–700) · risk ~4 times lower Childhood cancer: no threshold (stochastic), all periods thyroid (wk 3–4) iodine uptake (≥ wk 10–12) → I-131 contraindicated 0 2 8 15 25 40 Gestational period (weeks) Diagnostic NM: fetal dose usually <50 mGy → below the thresholds

Figure 10.6: Fetal radiation effects and threshold doses by gestational period.

Did this figure help you understand the topic?

10.4 Personal Dosimetry and Monitoring

10.4.1 Types of Personal Dosimeter1 questions

1. Film Badge:

–
Contains radiation-sensitive photographic film
–
Distinguishes radiation type and energy using filters of different thickness (copper, aluminium, lead, open window)
–
Measurement range: ~0.1 mSv – 10 Sv
–
Advantages: Permanent record, cheap, distinguishes radiation type
–
Disadvantages: No instant reading, sensitive to heat/humidity, requires monthly exchange, limited sensitivity at low doses
–
Exchange period: Usually monthly

2. TLD (Thermoluminescent Dosimeter):

–
Most common material: LiF (lithium fluoride)
–
Radiation energy is trapped in the crystal structure → released as light when heated
–
The amount of light is proportional to the absorbed dose
–
Measurement range: ~0.1 mSv – 10 Sv
–
Advantages: Small size, tissue-equivalent (LiF), reusable, good resistance to heat/humidity
–
Disadvantages: No instant reading, information is erased after reading (single readout), requires calibration
–
Exchange period: Usually monthly or quarterly

3. OSL (Optically Stimulated Luminescence) Dosimeter:

–
Most common material: Al₂O₃:C (carbon-doped aluminium oxide)
–
Similar to TLD but stimulated with laser light instead of heat
–
Advantages: Can be read multiple times (information is not erased), high sensitivity, small size, wide dose range
–
Disadvantages: No instant reading
–
Has largely replaced the film badge today
–
Exchange period: Usually monthly or quarterly

4. Pocket Ionisation Chamber (Pocket Dosimeter):

–
A small, pen-shaped ionisation chamber
–
Can be read instantly (self-reading)
–
Measurement range: 0 – 200 mR (typical)
–
Advantages: Instant reading, portable
–
Disadvantages: Limited dose range, sensitive to mechanical shock (may reset), no permanent record, low sensitivity

5. Electronic Personal Dosimeter (EPD):

–
Based on a semiconductor detector (silicon diode)
–
Instant display of dose and dose rate on a digital screen
–
Alarm function (when a preset dose or dose rate is exceeded)
–
Advantages: Instant reading, alarm, high sensitivity, digital record
–
Disadvantages: Expensive, requires a battery, energy dependence
–
Used as an additional dosimeter in high-dose environments (RAI therapy room, emergency response)
BOARD TIP
▸
"Dosimeter that gives an instant dose reading?"

→ Pocket ionisation chamber or electronic personal dosimeter (EPD).

▸
"Passive dosimeter that can be read multiple times?"

→ OSL.

▸
"Tissue-equivalent material?"

→ LiF (TLD).

▸
"The most important disadvantage of the film badge?"

→ No instant reading and sensitivity to heat/humidity.

10.4.2 Rules for Wearing Dosimeters3 questions

Use of a single dosimeter:

•
If no lead apron is used: On the front of the trunk (chest or waist level)
•
If a lead apron is used: Outside the apron, at collar (neck) level

Use of two dosimeters (recommended):

•
Collar dosimeter: Outside the apron, at neck level → reflects eye lens and thyroid dose
•
Waist dosimeter: Under the apron, at waist level → reflects trunk dose
•
Effective dose equivalent (HE) estimate, two-dosimeter formula (NCRP 122; the method accepted by the NRC in the USA):

HE ≈ 1.5 × Hwaist + 0.04 × Hcollar

For effective dose (E), same report: E ≈ 0.5 × Hwaist + 0.025 × Hcollar

ICRP 139 (2018), simple algorithm:

E ≈ Hwaist + 0.1 × Hcollar (0.05 × Hcollar with a thyroid shield)

WHY IT MATTERS · Effective dose equivalent (HE) versus effective dose (E)

•
What it is: The effective dose equivalent (HE) is the quantity defined in ICRP 26 (1977); the effective dose (E) replaced it in ICRP 60 (1991) and was updated in ICRP 103 (2007). The logic is the same (organ equivalent doses summed with tissue weighting factors), but the list of organs and the wT values differ.
•
In numbers: wT: gonads 0.25 (ICRP 26) → 0.20 (ICRP 60) → 0.08 (ICRP 103); breast 0.15 → 0.05 → 0.12; thyroid 0.03 → 0.05 → 0.04. From the same dosemeter readings, the NCRP 122 formulas give 1.5 × Hwaist + 0.04 × Hcollar for HE and 0.5 × Hwaist + 0.025 × Hcollar for E.
•
Why it matters clinically: Older radiopharmaceutical dose tables (ICRP 53, 1987) give HE; ICRP 80 (1998) and ICRP 128 (2015) give E. Values of the two quantities, or values calculated with different ICRP versions, are not interchangeable; when comparing, state which weights were used. US NRC regulations (10 CFR 20) still use the ICRP 26-based total effective dose equivalent (TEDE); the EU Basic Safety Standards use the effective dose based on ICRP 103.
•
In the exam: "The difference between HE and E?" → Same logic, different ICRP recommendations (26 versus 60/103) and different wT; not interchangeable.
•
Sources: ICRP 26 (1977), ICRP 60 (1991), ICRP 103 (2007); NCRP 122 (1995); US NRC 10 CFR 20.

Extremity dosimeter:

•
To measure hand dose during radiopharmaceutical preparation and injection
•
TLD ring dosimeter (usually on the dominant hand, palm side)
•
Limit: 500 mSv/year

General rules:

•
A dosimeter is personal → it cannot be given to someone else
•
Control dosimeter: Stored in a place not exposed to radiation → for background correction
•
A lost or damaged dosimeter must be reported immediately
•
Records should be kept for at least 30 years and until the worker reaches at least 75 years of age (IAEA recommendation)
BOARD TIP
▸
"Where is the dosimeter worn if a lead apron is used?"

→ OUTSIDE the apron, at collar level (if a single dosimeter). With two dosimeters: collar = outside the apron, waist = under the apron.

▸
"When is an extremity dosimeter needed?"

→ During radiopharmaceutical preparation/injection.

lead apron Collar dosimeter: OUTSIDE the apron, neck → eye lens + thyroid Waist dosimeter: UNDER the apron → trunk dose Ring dosimeter (TLD): dominant hand, palm side → extremity (limit 500 mSv/year) Single dosimeter: • no apron: front of trunk • apron: outside it, collar Two dosimeters (NCRP 122): HE ≈ 1.5 × Hwaist + 0.04 × Hcollar E ≈ 0.5 × Hwaist + 0.025 × Hcollar HE: effective dose equivalent Dosimeters are personal; the control dosimeter is kept in a radiation-free area.

Figure 10.7: Placement of collar, waist and ring dosimeters when a lead apron is used (schematic).

Did this figure help you understand the topic?

10.4.3 Biological Dosimetry3 questions

When physical dosimeters cannot be used or are insufficient (e.g. a radiation accident), dose can be estimated by biological methods:

1. Dicentric Chromosome Analysis (Gold Standard):

–
Counting of dicentric chromosomes in peripheral blood lymphocytes
–
A radiation-specific chromosome aberration → high specificity
–
Irradiation in G1 phase leads to chromosome-type aberrations, in S/G2 phase to chromatid-type aberrations. Dicentrics and ring chromosomes are unstable (lost as cells divide), whereas translocations are stable (past exposures are detected by FISH). At low LET the dicentric frequency follows αD + βD²
–
The dose–response curve is well defined
–
Disadvantage: Time-consuming (48-72 hours of culture), requires an expert
–
Dose range: 0.1 – 5 Gy

2. Lymphocyte Count (Rapid Triage):

–
After acute radiation exposure, the rate of fall in lymphocyte count is proportional to dose
–
Absolute lymphocyte count within 48 hours:
–
>1500/µL → <1 Gy (mild)
–
1000-1500/µL → 1-2 Gy (moderate)
–
500-1000/µL → 2-4 Gy (severe)
–
<500/µL → >4 Gy (very severe/lethal)
–
Advantage: Fast, easily available
–
Disadvantage: Low sensitivity; lymphopenia may have other causes

3. Micronucleus Test:

–
Faster and easier than dicentric analysis
–
Can be automated
–
Lower specificity

4. Electron Spin Resonance (ESR) – Tooth Enamel:

–
Measures radiation-induced free radicals in tooth enamel
–
Used to estimate chronic/cumulative dose
–
Retrospective dosimetry (can be measured even years later)

10.5 Radiation Safety Practice

10.5.1 The Three Basic Principles of Radiation Protection2 questions

1. Time:

–
Minimise the time spent near the radiation source
–
Dose = Dose rate × Time → if the time is halved, the dose is halved
–
Practical applications: Planning procedures in advance, practising, working quickly

2. Distance:

–
Reduce dose by moving away from the source
–
Inverse square law: Dose rate is inversely proportional to the square of the distance

Ḋ₁ × d₁² = Ḋ₂ × d₂²

•
If the distance is increased 2-fold → dose decreases 4-fold
•
If the distance is increased 3-fold → dose decreases 9-fold
•
Practical applications: Long forceps/tongs, remote injection systems, syringe shields

3. Shielding:

–
Place suitable shielding material between the source and the person
–
Gamma/X-rays: Lead (Pb), concrete, steel
–
PET (511 keV): Syringe shields are mostly made of tungsten; because of its high density, effective protection is achieved with a thinner shield
–
Beta: Low-Z materials (Perspex, acrylic) → first low Z, then high Z (to prevent bremsstrahlung)
–
Alpha: Paper, skin → no shielding needed (no external exposure risk)
–
Neutrons: Hydrogen-containing materials (water, paraffin, polyethylene) + boron/cadmium
BOARD TIP
▸
"Why is Perspex used first and lead second to shield a beta source?"

→ When beta particles are stopped in a high-Z material, bremsstrahlung (braking radiation) is produced → beta is first stopped with low Z, and the resulting bremsstrahlung is then shielded with lead. In the exam, "Shielding for P-32?" → Perspex (NOT lead!).

How many photons from a source reach a worker depends on distance, time and lead shielding; the shielding order and bremsstrahlung for a beta source are shown.

Continuous stream · no seeking

Key point: dose rate falls with the square of distance, dose rises in proportion to time, and each HVL of lead halves the photons passing; a beta source is shielded first with low-Z material.

Source: book (Γ Tc-99m 0.78, I-131 2.17, F-18 5.7 R·cm²/mCi·h; lead HVL 0.3 / 3 / 4 mm); example 20 mCi (740 MBq) point source, body attenuation and scatter not included. Beta step: P-32 Emax 1.711 MeV (ENSDF), range NIST ESTAR, bremsstrahlung fraction Cherry, Sorenson & Phelps (Physics in Nuclear Medicine). Magnifier: Rutherford scattering; bremsstrahlung spectrum and angular distribution after Bethe–Heitler (Attix, Introduction to Radiological Physics and Radiation Dosimetry); atomic radii not to scale.

How to use: press and drag the worker (release and it moves back and forth on its own); hold ← → to change the distance smoothly; one click on the drawing pauses and resumes, double-click enlarges.

Speed
Test yourself:Chapter 10 · question 15Chapter 10 · question 7
Did this animation help you understand the topic?

Figure 10.8: The three principles of radiation protection: time, distance (inverse square law) and the correct shielding order for a beta source.

10.5.2 Components of a Radiation Safety Programme

Radiation Safety Committee (RSC):

•
Sets institutional radiation safety policies
•
Members: Radiation safety officer (RSO), management representative, representatives of departments using radiation
•
Meets at least once a year

Radiation Safety Officer (RSO):

•
Responsible for day-to-day radiation safety operations
•
Oversees compliance with licence conditions
•
Staff training, dosimetry follow-up, contamination control
•
Radioactive material inventory and waste management

Staff Training:

•
Initial training on starting work and annual refresher training
•
Basics of radiation physics, protection principles, emergency procedures
•
Information on the rights and risks of pregnant workers

10.5.3 Security of Radioactive Material2 questions

Receipt and recording of radioactive material:

•
Wipe test on receipt of a package (contamination check)
•
Dose rate measurement at the package surface and at 1 metre
•
Inventory record: Radionuclide, activity, date, intended use

Transport classification (DOT/IAEA):

CategoryPackage SurfaceAt 1 mLabel Colour
White-I≤0.005 mSv/h–White
Yellow-II≤0.5 mSv/h≤0.01 mSv/hYellow
Yellow-III≤2 mSv/h≤0.1 mSv/hYellow

Transport Index (TI):

•
The dose rate at 1 metre expressed in mrem/h
•
White-I: TI = 0
•
Yellow-II: TI ≤ 1.0
•
Yellow-III: TI ≤ 10.0

Storage of radioactive material:

•
In a locked, labelled, ventilated area
•
Radiation warning sign (trefoil symbol) in a visible place
•
Different radionuclides should be appropriately segregated
•
Volatile materials should be stored in a fume hood
•
Radioiodine solutions are volatile (I₂ vapour): Oxidation and acidic pH increase volatility. NaI solution is therefore stored at alkaline pH with a reducing agent added (e.g. thiosulfate) and opened in a fume hood; capsule form is preferred where possible
BOARD TIP
▸
"0.3 mSv/h at the package surface, 0.005 mSv/h at 1 m → which category?"

→ Yellow-II (surface ≤0.5, 1m ≤0.01). Transport Index = dose rate at 1 m (mrem/h) → 0.005 mSv/h = 0.5 mrem/h → TI = 0.5. Package classification questions come up frequently in the exam!

10.5.4 Area Classification6 questions

Area TypeDose Rate CriterionSignage
Unrestricted AreaPublic limit (1 mSv/year)No sign needed
Restricted AreaControlled accessRadiation warning sign
Radiation Area0.05 mSv/h (5 mrem/h) @ 30 cm"Caution: Radiation Area"
High Radiation Area1 mSv/h (100 mrem/h) @ 30 cm"Caution/Danger: High Radiation Area"
Very High Radiation Area5 Gy/h (500 rad/h) @ 1 m"Grave Danger: Very High Radiation Area"

Important rules:

•
High Radiation Area: Access control is required (lock, alarm or continuous surveillance)
•
Very High Radiation Area: A lethal dose may be received within minutes → additional security measures are mandatory
•
Areas where radioactive material is used/stored: "Caution: Radioactive Materials" sign

The European Basic Safety Standards (BSS) approach:

•
International (IAEA) and regional (EU) Basic Safety Standards have been published on the basis of ICRP recommendations (the most recent from 2007, ICRP 103); the EU BSS Directive has been transposed into the legislation of all member states
•
Work areas: Controlled areas, where access and supervision are stricter, and supervised areas, where the risk is lower
•
Worker categories: Category A workers are those likely to receive an effective dose above 6 mSv per year, or an equivalent dose above 15 mSv/year to the eye lens or 150 mSv/year to the skin/extremities. Category B workers are those likely to exceed the dose limits set for the public but who do not fall into Category A
•
Institutional local rules may be stricter than national and international regulations; local rules must always be followed
•
Contamination and intake of radioactive material must be prevented in particular, because they can render time–distance–shielding measures ineffective

10.5.5 Specific Gamma-Ray Constant (Γ – Exposure Rate Constant)2 questions

Definition: Gives the exposure rate of a radionuclide per unit activity at unit distance.

Formula:

Ẋ = Γ × A / d²

•
Ẋ: Exposure rate (R/h or mR/h)
•
Γ: Specific gamma-ray constant (R·cm²/mCi·h)
•
A: Activity (mCi)
•
d: Distance (cm)

Γ values of common radionuclides:

RadionuclideΓ (R·cm²/mCi·h)
Tc-99m0.78
I-1312.17
I-1231.55
Ga-670.76
Tl-2010.45
F-185.7
Co-6013.2
Cs-1373.26
Ra-2268.25
BOARD TIP
▸
The Γ value of F-18 (5.7) is much higher than that of Tc-99m (0.78) → more shielding is needed when preparing PET radiopharmaceuticals. The Γ value of I-131 (2.17) is also high → patients receiving RAI therapy emit significant radiation into their surroundings. For exam questions such as "which radionuclide poses a greater external radiation risk?", compare the Γ values.

10.6 Contamination and Decontamination

10.6.1 Types of Contamination

Fixed Contamination:

•
Radioactive material firmly bound to a surface
•
Cannot be removed by a wipe test
•
Poses an external radiation risk but a low risk of spread

Removable Contamination:

•
Radioactive material that is easily detached from a surface
•
Detected by a wipe test
•
High risk of spread and internal contamination → more dangerous
•
NRC: ≤200 dpm/100 cm² (I-131, In-111, etc.) / ≤2000 dpm/100 cm² (Tc-99m, Ga-67, Tl-201) → unrestricted areas

10.6.2 Detection of Contamination1 questions

Wipe Test (Smear Test):

•
An area of 100 cm² is wiped with filter paper or a cotton swab
•
Counted in a well counter
•
Measures removable contamination
•
Frequency: Weekly in areas where radioactive material is used
•
Note: Mo-99 breakthrough is not a wipe test; it is measured in a lead pot in the dose calibrator

Survey Meter Monitoring:

•
Surface survey with a GM counter or ionisation chamber
•
Detects both fixed and removable contamination
•
Measures total contamination (cannot distinguish)
•
The work area should be surveyed after every use

10.6.3 Decontamination Procedures2 questions

Skin Decontamination:

1. Gently wash the contaminated area with warm water and soap (3-4 minutes) 2. Check with a survey meter after washing 3. Repeat if necessary (maximum 3-4 times) 4. DO NOT scrub hard → if skin integrity is broken, radioactive material is absorbed 5. For resistant contamination: Potassium permanganate or DTPA solution 6. For wound contamination: Copious irrigation with water → surgical debridement as a last resort

Surface Decontamination:

1. Mark and confine the contaminated area (surround with absorbent paper) 2. Wipe from the outside inwards (to prevent spread of contamination) 3. Clean with a suitable detergent/solvent 4. Check with a survey meter after each cleaning 5. If background level cannot be reached → mark the area and leave it to decay (for short-lived isotopes)

Procedure for a Major Contamination Event (Spill):

Minor spill:

1. Put on gloves, cover the contaminated area with absorbent paper 2. Clean from the outside inwards 3. Check with a survey meter 4. Report to the RSO

Major spill:

1. Leave the area immediately (do not spread the contamination) 2. Lock/mark the room → prevent others from entering 3. Report to the RSO immediately 4. Decontaminate contaminated persons 5. Clean-up by a professional decontamination team

On-Site and Hospital Response in a Radiation Accident

•
Triage has two steps: first, casualties are sorted by the severity of their medical condition using conventional triage systems (not by radiation exposure or contamination criteria); in the second step, people who may have been exposed or contaminated are identified. Life-threatening conditions always take priority over radiological concerns
•
Radiation does not cause an immediately life-threatening condition; unconsciousness or immediately visible signs of injury are rarely explained by radiation, and if present another cause is sought
•
At the scene: a patient with life-threatening injuries is transported to hospital without waiting for a contamination survey; the hospital is told about the conventional injuries and any possible exposure or contamination. Wounds are covered with sterile dressings. The radiological survey must not delay medical care or transport
•
Clothing suspected to be contaminated is removed (unless medically contraindicated), placed in plastic bags and labelled; clothing can serve as a sample for identifying the radionuclide
•
Radiological triage: areas measuring more than 2-3 times background are considered contaminated. The alpha background is taken as zero; if an alpha emitter is detected at any level, the person is considered contaminated. When internal contamination is suspected, nasal swabs are taken within about 60 minutes of intake; a negative oral swab means little
•
At the hospital: the hospital emergency plan is activated; an ambulance reception area and a treatment area are set up, access is restricted, survey instruments are checked and background levels recorded; if time permits the floor is covered (emergency care is not delayed for this). The ambulance crew stays with the vehicle until they, the vehicle and the equipment have been surveyed and released
•
A contaminated patient is not a significant source of external exposure for the team; the main risk is the spread of contamination to the premises, equipment and other people
•
Order of decontamination: wounds → body orifices (eyes, ears, mouth, nose) → highly contaminated skin → less contaminated skin. Wounds are irrigated with saline or water and monitored after each irrigation; irrigation with DTPA may be considered for Am, Pu and Cm; embedded radioactive foreign bodies are removed and kept for analysis. Internal contamination is assumed in a contaminated wound. Hair is washed with shampoo; the skin is not shaved
•
Uninjured people with external contamination only are decontaminated at a facility other than the hospital to conserve hospital resources; in an event affecting many people, radiological triage of the worried well is performed away from the treatment area
•
When high-dose whole-body irradiation is suspected: an immediate full blood count with differential, followed by an absolute lymphocyte count every 6 hours for 48 hours (see 10.4.3)
BOARD TIP
▸
"The most important rule in skin decontamination?"

→ DO NOT scrub hard → skin damage increases absorption of radioactive material.

▸
"Direction of wiping in surface decontamination?"

→ From the outside inwards (to prevent spread of contamination).

▸
"First step in a major spill?"

→ Leave the area and report to the RSO.

BOARD TIP
▸
"Priority in a radiation accident patient who is both seriously injured and contaminated?"

→ Treatment of the life-threatening condition and stabilisation; decontamination and radiological survey must not delay it.

▸
"Order of decontamination in a patient?"

→ Wounds → body orifices → highly contaminated skin → less contaminated skin.

▸
"Criterion for 'contaminated' in radiological triage?"

→ A reading above 2-3 times background; any level of an alpha emitter.

Radioactive spill Minor spill 1 Put on gloves; cover the area with absorbent paper 2 Clean from the outside inwards 3 Check with a survey meter 4 Report to the RSO Major spill 1 Leave the area at once (do not spread contamination) 2 Lock / mark the room 3 Report to the RSO IMMEDIATELY 4 Decontaminate contaminated persons 5 Clean-up by a professional team Wiping direction: outside inwards

Figure 10.9: What to do in a minor and a major radioactive spill.

Did this figure help you understand the topic?

10.6.4 Internal Contamination and Its Treatment2 questions

Internal contamination occurs when radioactive material is taken into the body (inhalation, ingestion, through a wound).

Treatment approaches:

RadionuclideTreatmentMechanism
I-131Potassium iodide (KI)Thyroid blockade (dilution with stable iodine)
Cs-137Prussian blueBinds in the GI tract and increases faecal excretion
Plutonium, AmericiumDTPA (Ca-DTPA or Zn-DTPA)Chelation → renal excretion
Tritium (H-3)Plenty of water (forced diuresis)Shortens the biological half-life
Strontium (Sr-90)Aluminium phosphate, barium sulfateReduces GI absorption
UraniumSodium bicarbonateReduces renal tubular damage
BOARD TIP
▸
Match the treatments for internal contamination: I-131 → KI, Cs-137 → Prussian blue, Plutonium → DTPA, Tritium → plenty of water. In the exam, "treatment of Cs-137 contamination after a nuclear accident?" → Prussian blue.
▸
"When should KI be given for thyroid protection?"

→ Before exposure or within 4 hours at the latest (effectiveness decreases with time).

10.7 Radiation Safety in Pregnancy and Breastfeeding

10.7.1 Diagnostic Nuclear Medicine in Pregnancy2 questions

General principles:

•
Pregnancy is not an absolute contraindication to nuclear medicine studies
•
A risk–benefit analysis should be performed
•
If possible, the study should be postponed until after delivery
•
If postponement is not possible, the protocol that gives the most diagnostic information at the lowest dose should be chosen

Estimated fetal doses in common studies:

StudyRadiopharmaceuticalTypical ActivityEstimated Fetal Dose
Lung perfusionTc-99m MAA2-4 mCi0.2-0.6 mGy
Bone scintigraphyTc-99m MDP20-25 mCi4-5 mGy
Renal scintigraphyTc-99m MAG310 mCi~5-7 mGy
Thyroid scintigraphyTc-99m pertechnetate5-10 mCi~2-4 mGy
V/Q scintigraphyTc-99m MAA + Xe-133Combined0.5-1.0 mGy
FDG PET/CTF-18 FDG10-15 mCi10-20 mGy

Important rules:

•
In diagnostic nuclear medicine studies the fetal dose is usually <50 mGy → below the threshold for deterministic effects
•
Termination is not recommended below 50 mGy (ACR); ICRP 84: it cannot be justified at <100 mGy
•
Above 100 mGy, detailed risk assessment and counselling are needed
•
Hydration and frequent bladder emptying → reduce fetal dose (especially the uterine dose from the adjacent bladder)
•
I-131 is absolutely contraindicated in pregnancy → it crosses the placenta, and the fetal thyroid concentrates iodine from week 10-12 onwards → fetal hypothyroidism/thyroid ablation
BOARD TIP
▸
"Radiopharmaceutical absolutely contraindicated in pregnancy?"

→ I-131 (and all forms of radioiodine → fetal thyroid damage).

▸
"The safest lung perfusion agent in pregnancy?"

→ Tc-99m MAA (can be performed at a low dose, fetal dose <1 mGy).

▸
"Dose threshold for termination of pregnancy?"

→ Not recommended below 50 mGy.

10.7.2 Radioiodine Exposure in Pregnancy

Fetal thyroid development and iodine uptake:

•
The fetal thyroid begins to develop at ~week 3-4
•
It begins to concentrate iodine from week 10-12 onwards
•
In the 2nd and 3rd trimesters the fetal thyroid concentrates iodine at a higher rate than the adult thyroid

Consequences of I-131 exposure:

•
1st trimester (<10 weeks): The fetal thyroid does not yet take up iodine → low risk of thyroid damage, but there is a whole-body dose risk
•
2nd-3rd trimester (>10 weeks): The fetal thyroid concentrates iodine → fetal hypothyroidism, thyroid ablation, congenital hypothyroidism
•
At high doses: Risk of fetal thyroid cancer (theoretical)

If I-131 is given inadvertently:

•
Give KI (potassium iodide) immediately → fetal thyroid blockade
•
Effectiveness depends on the time since exposure (the earlier the better)
•
Hydration and frequent voiding → reduce maternal and fetal dose
•
Dosimetric assessment and genetic counselling

10.7.3 Radiation Safety in Breastfeeding3 questions

Radiopharmaceuticals can pass into breast milk → they may deliver a radiation dose to the breastfed infant. The duration of breastfeeding interruption varies with the radiopharmaceutical.

Durations of breastfeeding interruption:

RadiopharmaceuticalBreastfeeding Interruption
Tc-99m pertechnetate12 hours
Tc-99m MAA12 hours
Tc-99m MDPNot needed (EANM: ≥4 hours)
Tc-99m MIBINot needed (EANM: ≥4 hours)
Tc-99m DTPANot needed (EANM: ≥4 hours)
Tc-99m MAG3Not needed (EANM: ≥4 hours)
Tc-99m DMSANot needed (EANM: ≥4 hours)
Tc-99m sulphur colloidNot needed (EANM: ≥4 hours)
Tc-99m RBC (labelled red cells)12 hours (in vivo labelling)
I-123 (NaI)3-4 days (older recommendation: ≥3 weeks)
I-123 MIBG≥24 hours (EANM)
I-131 (any dose)Stop completely (permanent)
Ga-67 citrate≥3-4 weeks (stopped in practice)
Tl-20148 hours
F-18 FDGNot needed (very little passes into milk); close contact with the infant is limited for the first ~4 hours (EANM/SNMMI 2024). The US ACMUI recommends a 4-hour interruption
In-111Not needed (octreotide, leucocytes)

Try it in the calculator: Breastfeeding interruption →

Important rules:

•
I-131: Breastfeeding must be stopped permanently (it may be resumed in a subsequent pregnancy) → passes into milk at a high rate and concentrates in the infant's thyroid
•
During the interruption, milk should be expressed and discarded (to maintain milk production – except for I-131)
•
The duration of close contact with the infant should also be limited (external radiation)
•
Most Tc-99m compounds do not require interruption; EANM recommends 12 hours for pertechnetate and MAA and ≥4 hours for the others
•
Ga-67 requires a long interruption because of its long half-life and high transfer into milk
BOARD TIP
▸
"When can breastfeeding be resumed after I-131?"

→ IT CANNOT → permanent interruption (breastfeeding of that infant is stopped completely).

▸
"Breastfeeding interruption after Tc-99m pertechnetate?"

→ 12 hours.

▸
"Diagnostic agent requiring the longest breastfeeding interruption?"

→ Apart from I-131, Ga-67 (≥3-4 weeks; stopped in practice). The I-131 and breastfeeding question is a classic in the exam!

1 h 4 h 12 h 1 day 2 days 1 week 4 weeks 6 months Breastfeeding interruption (logarithmic scale) Tc-99m MDP, MIBI, DTPA, MAG3, DMSA, sulphur colloid not needed (EANM: ≥4 hours) F-18 FDG not needed; close contact limited for first 4 h Tc-99m pertechnetate 12 hours Tc-99m MAA 12 hours Tc-99m RBC (in vivo) 12 hours I-123 MIBG ≥24 hours (EANM) Tl-201 48 hours I-123 (NaI) 3–4 days Ga-67 citrate ≥3–4 wk (stopped) I-131 (any dose) PERMANENT STOP In-111 (octreotide, leucocytes): not needed Light end: upper limit of range. Milk is expressed and discarded (after I-131 breastfeeding ends).

Figure 10.10: Durations of breastfeeding interruption by radiopharmaceutical.

Did this figure help you understand the topic?

10.7.4 Pregnancy Screening Before a Nuclear Medicine Study1 questions

Indications for a pregnancy test:

•
All women of childbearing age (usually 12-50 years)
•
Regardless of the last menstrual period (early pregnancy may be missed)
•
The old '10-day rule', which restricted non-urgent studies in women of childbearing age to the first 10 days of the menstrual cycle, is no longer applied strictly; what matters is asking about pregnancy and checking β-hCG when necessary
•
Mandatory especially before I-131 therapy

Practice:

•
Serum or urine β-hCG test
•
If the result is negative, the study can be performed
•
If the result is positive → risk–benefit analysis, consideration of an alternative study
•
I-131 therapy → absolutely not given if the pregnancy test is positive

10.8 Patient Release Criteria (After RAI Therapy)

10.8.1 NRC Release Criteria (10 CFR 35.75)2 questions

•
The US Nuclear Regulatory Commission (NRC) demonstrates the release criterion (≤5 mSv to others) in three ways:

Criterion 1 – Activity-Based:

•
The patient may be released when the retained activity is ≤33 mCi (1.22 GBq)
•
Simple and quick assessment
•
Usually used for I-131 therapy

Criterion 2 – Dose-Rate-Based:

•
The patient may be released when the dose rate at 1 metre is ≤7 mR/h (0.07 mSv/h)
•
Measured with a survey meter

Criterion 3 – Patient-Specific Dosimetric Calculation:

•
The total dose that individuals around the patient will receive is calculated according to the individual patient's circumstances
•
The total dose to any individual around the patient must be ≤5 mSv (500 mrem)
•
This method is the most flexible → release may be possible even with higher activities
•
The patient's living conditions, contact times and distances are taken into account

Comparison:

CriterionParameterValueAdvantage
Activity-basedRetained activity≤33 mCi (1.22 GBq)Simple, quick
Dose-rate-basedDose rate at 1 m≤7 mR/h (0.07 mSv/h)Based on measurement
Patient-specificDose to others≤5 mSv (total)Most flexible

Try it in the calculator: I-131 release →

BOARD TIP
▸
"Release criterion after I-131 therapy?"

→ 3 options: ≤33 mCi retained activity OR ≤7 mR/h @ 1 m OR ≤5 mSv by patient-specific calculation. The values asked most often in the exam: 33 mCi and 7 mR/h.

▸
"Dose limit to others for a child/pregnant woman?"

→ 1 mSv (5 mSv for an adult).

Release after I-131 therapy (NRC 10 CFR 35.75) Criterion 1 – Activity Retained activity ≤1.22 GBq (33 mCi) simple, quick OR Criterion 2 – Dose rate dose rate at 1 m ≤0.07 mSv/h (7 mR/h) measurement-based (survey meter) OR Criterion 3 – Patient-specific Any nearby individual: dose ≤5 mSv most flexible (living conditions, contact time, distance) Any one is sufficient → release + written instructions. Dose constraint for child/pregnant woman 1 mSv (ICRP 94, IAEA).

Figure 10.11: NRC release criteria after I-131 therapy.

Did this figure help you understand the topic?

10.8.2 Patient Instructions After Release

Patients released after I-131 therapy should be given written instructions to reduce the radiation exposure of people around them:

General precautions:

•
Limit the time of close contact with other people (especially for the first 2-5 days)
•
Avoid contact with children and pregnant women in particular
•
1 mSv for a child/pregnant woman: ICRP 94/IAEA dose constraint (threshold for written instructions under the NRC)
•
Sleep in a separate bed (usually for the first 3-7 days or depending on the dose)
•
Minimise close contact such as kissing and hugging

Hygiene precautions:

•
Use a separate bathroom (if possible) or clean it after use
•
When using the toilet: Urinate sitting down (including men), flush twice, wash hands afterwards
•
Use separate cutlery, cups and towels (for the first few days)
•
Radioiodine is also excreted in saliva and sweat → keep personal items separate

Distance and time recommendations (for a typical I-131 dose of 100-200 mCi):

Contact PersonDistanceTime Restriction
Adult (spouse/partner)≥1 metreFirst 3-5 days
Children (<18 years)≥2 metresFirst 5-7 days
Pregnant women≥2 metresFirst 7 days or longer
Work colleagues≥1 metreReturn to work usually after 2-5 days

Travel restrictions:

•
Airport security detectors can detect radioactivity → the patient should carry an information card
•
Alarm period: May range from a few weeks to 3 months after I-131 therapy
•
Problems may arise at border crossings during international travel

Hydration and excretion:

•
Plenty of fluids → speeds renal excretion → reduces bladder dose
•
Frequent voiding
•
Lemon drops/sour sweets → reduce salivary gland dose (increase salivary flow)
•
Avoiding constipation → reduces bowel dose

10.8.3 Death of a Radioactive Patient

Special procedures apply when a patient containing radioactive material dies:

General rules:

•
The RSO must be informed immediately
•
If an autopsy is required: It should be performed under radiation safety precautions
•
Cremation: Should be postponed until the activity in the body falls below certain limits
•
Burial: Usually no restriction (soil provides sufficient shielding)
•
Staff preparing the body for the funeral: Should wear a dosimeter, gloves and protective clothing

NRC rules:

•
If a patient who has received I-131 therapy dies → the retained activity in the body should be assessed
•
For cremation, a retained activity of usually ≤30 mCi or a sufficient decay period should be awaited
•
Organ donation: Radioactive organs cannot be donated

10.8.4 Reporting of Misadministration2 questions

The NRC requires certain errors in radiopharmaceutical administration to be reported:

Events that must be reported (Medical Event – 10 CFR 35.3045):

•
Administration to the wrong patient
•
Administration of the wrong radiopharmaceutical
•
Wrong route of administration (oral instead of IV, etc.)
•
Deviation of more than ±20% from the prescribed dose (for therapeutic doses)
•
Dose delivered to an organ other than the intended target organ
•
Additional condition (not a criterion on its own; it must be present together with one of the errors above): an effective dose difference of >50 mSv from that prescribed OR an equivalent dose difference of >500 mSv in any organ or tissue

Reporting process:

1. Immediately: Report to the RSO 2. By the next calendar day: Report to the NRC (or the relevant regulatory authority) by telephone/fax 3. Within 15 days: Submit a written report 4. Patient notification: The patient (or legal representative) should be informed within 24 hours 5. Referring physician: The referring physician should be informed

BOARD TIP
▸
"Reporting time for a medical event?"

→ to the NRC by the next calendar day, written report within 15 days.

▸
"Acceptable deviation in a therapeutic dose?"

→ ±20% (outside this is a medical event).

▸
"Medical event criterion in diagnostic studies?"

→ Wrong patient, radiopharmaceutical or route BUT only if there is >50 mSv effective or >500 mSv organ dose → rare in diagnostic studies because of the low doses.

Key Points for the Board

Operational Quantities and EU BSS – Additional Points:

•
Effective dose cannot be measured → H*(10) (ambient), Hp(10) (whole body), Hp(0.07) (skin), Hp(3) (eye lens)
•
Controlled area vs supervised area
•
Category A: likely to receive >6 mSv/year effective, >15 mSv/year eye lens, >150 mSv/year skin/extremity; Category B: those who may exceed the public limit

Dosimetric Quantities:

•
Absorbed dose: Gray (Gy) = 100 rad → a physical quantity, does not take radiation type into account
•
Equivalent dose: Sievert (Sv) = 100 rem → corrected with the radiation weighting factor (wR)
•
Effective dose: Sievert (Sv) → corrected with the tissue weighting factor (wT) → indicator of stochastic risk
•
For photons and beta wR = 1 → absorbed dose (Gy) = equivalent dose (Sv) numerically
•
For alpha wR = 20 → 1 Gy of alpha = 20 Sv equivalent dose
•
Exposure (roentgen) is valid only for photons and in air
•
In air 1 R ≈ 0.87 rad; in bone the f factor (rad/R) rises markedly at low photon energies (photoelectric, Z³)
•
Effective dose per MBq: Tl-201 (~0.14-0.22) and Ga-67 (~0.10 mSv/MBq) are much higher than most Tc-99m agents (~0.005-0.01)

ICRP 103 – Tissue Weighting Factors (wT):

•
Highest wT (0.12): Bone marrow, colon, lung, stomach, breast, remainder tissues
•
Gonad wT: 0.20 in ICRP 60 → 0.08 in ICRP 103 (reduced)
•
Breast wT: 0.05 in ICRP 60 → 0.12 in ICRP 103 (raised)
•
Total wT = 1.00

MIRD Formalism:

•
Dose = Cumulated activity (Ã) × S value
•
Cumulated activity = area under the activity–time curve = total number of decays
•
Effective half-life: Teff = (Tphys × Tbiol)/(Tphys + Tbiol) → always shorter than both, closer to the shorter one
•
S value: A physical parameter depending on radionuclide emissions, organ geometry and mass

Dose Limits:

•
Occupational: 20 mSv/year (5-year average; ≤50 mSv in a single year)
•
Public: 1 mSv/year
•
Eye lens (ICRP 118): 20 mSv/year average (old: 150 mSv/year)
•
Skin/extremity: 500 mSv/year
•
Fetal: 1 mSv (ICRP) or 5 mSv (NRC) for the whole pregnancy
•
Dose limits do not apply to patients → managed by justification and optimisation

Personal Dosimetry:

•
TLD material: LiF (tissue-equivalent)
•
OSL material: Al₂O₃:C → can be read multiple times (information is not erased)
•
Film badge: No instant reading, sensitive to heat/humidity
•
Instant reading: Pocket ionisation chamber or electronic personal dosimeter (EPD)
•
Lead apron + single dosimeter → outside the apron, at collar level
•
Gold standard in biological dosimetry: Dicentric chromosome analysis
•
Dicentrics/rings are unstable (lost with division); past exposure → FISH analysis of stable translocations

Radiation Safety:

•
ALARA: As Low As Reasonably Achievable
•
The protection triad: Time, Distance, Shielding
•
Inverse square law: Distance 2× → dose decreases 4×
•
Beta shielding: Low Z (Perspex) → to prevent bremsstrahlung
•
Shielding for P-32 → Perspex (NOT lead)
•
PET (511 keV) syringe shields → tungsten (high density)
•
Radioiodine solution is volatile: acidic pH/oxidation increase volatility → alkaline pH + reducing agent, fume hood

Package Transport Categories:

•
White-I: Surface ≤0.005 mSv/h, TI = 0
•
Yellow-II: Surface ≤0.5 mSv/h, 1 m ≤0.01 mSv/h, TI ≤1.0
•
Yellow-III: Surface ≤2 mSv/h, 1 m ≤0.1 mSv/h, TI ≤10.0

Area Classification:

•
Radiation Area: 0.05 mSv/h @ 30 cm
•
High Radiation Area: 1 mSv/h @ 30 cm
•
Very High Radiation Area: 5 Gy/h @ 1 m

Contamination and Decontamination:

•
Removable contamination: ≤200 (I-131) / ≤2000 (Tc-99m) dpm/100 cm²
•
Skin decontamination: Warm water + soap, DO NOT scrub hard
•
Surface decontamination: Wipe from the outside inwards
•
Major spill: Leave the area → report to the RSO

Treatment of Internal Contamination:

•
I-131 → KI (potassium iodide)
•
Cs-137 → Prussian blue
•
Plutonium/Americium → DTPA
•
Tritium → Plenty of water (forced diuresis)

Pregnancy and Breastfeeding:

•
I-131 is absolutely contraindicated in pregnancy → fetal thyroid damage (from week 10-12 onwards)
•
Termination of pregnancy is not recommended below 50 mGy
•
Breastfeeding must be stopped permanently after I-131
•
Tc-99m compounds: Most require no interruption; 12 hours for pertechnetate and MAA
•
Ga-67: ≥3-4 weeks (stopped in practice), Tl-201: 48 hours

Release Criteria (After I-131 Therapy):

•
Activity-based: ≤33 mCi (1.22 GBq)
•
Dose-rate-based: ≤7 mR/h @ 1 m
•
Patient-specific: Dose to others ≤5 mSv (child/pregnant woman: ≤1 mSv ICRP/IAEA constraint)

Medical Event (Misadministration):

•
Deviation of more than ±20% in a therapeutic dose
•
Wrong patient, wrong radiopharmaceutical, wrong route of administration
•
Reporting: to the NRC by the next calendar day, written report within 15 days
•
The patient should be informed within 24 hours

Current guidelines on this chapter's topics

The latest guidelines and consensus documents on the procedures covered in this chapter (official EANM index and primary sources, November 2026). Comparison of the text with these versions is in progress; consult the guideline itself for details.

  1. 2024Generic Procedures for Medical Response During a Nuclear or Radiological Emergency (IAEA EPR-Medical) (opens in a new tab)

Finished the chapter?

This chapter has 24 questions. Solve them now to consolidate the topic.

Solve questions →
Previous chapterImage Quality and ArtefactsNext chapterRadionuclide Production and Generators

You can also swipe sideways to change chapter

Step 1/3

Physical, biological and effective half-life

The radiopharmaceutical in the organ decreases in two ways: physical decay (the atom flashes and fades in place) and biological excretion (the atom leaves the organ). Because both act together, the organ activity falls faster than with either alone: 1/Teff = 1/Tphys + 1/Tbiol. The effective half-life is always shorter than both and closer to the shorter one (6 h and 24 h → 4.8 h).

OrganInside 80Decayed 0Excreted 0t = 0 h
025507510006121824 hIn organ (%)Teff = 4.8 hPhysical 6.0 · Biological 24.0 h
PhysicalBiologicalEffective

Example from the book: Tphys = 6 h, Tbiol = 24 h → Teff = (6 × 24) / (6 + 24) = 4.8 h. In the cumulated activity step A₀ = 100 MBq and organ fraction f = 1 (schematic). Atoms are not to scale; each atom either decays physically or is excreted from the organ.

Step 1/5

Distance: inverse square law

Photons spread out from the source in all directions; with distance the same number of photons covers a larger area. Dose rate is inversely proportional to the square of distance: Ḋ₁·d₁² = Ḋ₂·d₂². Doubling the distance cuts dose 4-fold, tripling it 9-fold. That is why long tongs and remote injection work.

Ẋ = Γ·A/d² = 6.2 mR/hPhotons reaching the body/s: 0.0Worker50100200d = 50 cmTc-99m20 mCi

Example: 20 mCi (740 MBq) source. Γ (R·cm²/mCi·h) and lead HVL from the book: Tc-99m 0.78 / 0.3 mm, I-131 2.17 / 3 mm, F-18 5.7 / 4 mm. Ẋ = Γ·A/d² applies to a point source; body attenuation and scatter are not included. Photons are generated in random 3D directions; hits on the body fall as 1/d².