Contamination survey and wipe test (work area, personnel, incoming packages)
Awaiting physician review
To find, measure and record residual radioactive contamination on work areas, staff and incoming packages in good time.
What you need
- Surface contamination monitor checked daily with a check source (GM or scintillation probe)
- Dose rate meter (ion chamber)
- Dry filter paper or wipes, numbered tubes
- Well counter (with a background count)
- Gloves, lab coat, record form and area map
Steps
- Check the monitor's battery and its response to the check source; start in a low-background area and record the background count.
- Area survey: work bench, inside the fume hood, dose calibrator, injection and imaging areas, floor, around waste bins; move the probe slowly and close to the surface.
- Wipe test: at each point wipe about 100 cm² with dry paper using moderate pressure; put the paper in its tube and mark the point on the area map.
- Count the samples with a background sample in the well counter; convert count rate to dpm/100 cm² using the efficiency and compare with the institution's threshold.
- Staff: hands, shoes and clothing are monitored on leaving the controlled area; if contaminated, clothing is removed and skin is washed immediately.
- Incoming package: check a labelled package within 3 hours during working hours; with gloves, look for damage, wetness or crushing; measure dose rate at the surface and at 1 m and compare with the transport index on the label.
- Wipe the outer surface of the package (about 300 cm²) and count it; compare the inner vial with the packing slip and look for breakage or leakage; survey the empty packaging and, if clean, remove or deface its radiation labels.
- Clean any point above the threshold and re-survey; record every result with date, location and the person measuring; report any unexplained high value to the radiation protection officer.
Key points
- A wipe test measures only removable contamination; fixed contamination needs direct probe survey.
- The NRC model procedure (NUREG-1556 Vol. 9, 2019) gives example removable contamination levels for restricted areas: 20 000 dpm/100 cm² for Tc-99m, Tl-201, Ga-67; 2 000 for I-131, In-111, Lu-177. These are examples of institutional levels; the institution's and national thresholds apply.
- The same source gives example frequencies: weekly for use, storage and waste areas; monthly for laboratories using only small quantities. IAEA SSG-46 requires periodic monitoring of controlled and supervised areas, and contamination and dose rate measurements on a damaged incoming package.
- A pancake GM probe over-reads dose rate for most radionuclides; an ion chamber is preferred for dose rate.
- In the US package checks follow 10 CFR 20.1906 (3 hours during working hours; the first 3 hours of the next working day otherwise). Türkiye: the NDK guide on the radiation protection programme in nuclear medicine (RSGD-KLV-030, 2019) treats twice background or more as contamination; wipe test limits are 0.001 µCi/cm² (37 Bq/cm²) for I-131 and 0.01 µCi/cm² (370 Bq/cm²) for Tc-99m.
Common mistakes
- Using the monitor without a check-source test
- Not taking a background count
- Moving the probe fast and far from the surface
- Leaving the incoming package unchecked
- Discarding clean packaging without removing its labels
Safety. Wear gloves; mark a contaminated point and keep it out of use until cleaned; monitor the instrument and yourself at the end of work.
Sources: NDK RSGD-KLV-030 Nükleer tıp uygulamalarında radyasyondan korunma programına ilişkin kılavuz (2019) · IAEA SSG-46 (2018), paras 4.113-4.116 · NUREG-1556 Vol. 9 Rev. 3 (NRC 2019), Appendix R · 10 CFR 20.1906 (eCFR, 2026) · Pennsylvania Üniversitesi nükleer tıp kullanıcı kılavuzu (2026), paket kabulü ve silme testi. Last reviewed: 23 November 2026.