Patient procedures · 7 of 7

Patient positioning: bone scintigraphy, myocardial perfusion SPECT and PET/CT

Awaiting physician review

To position the patient comfortably and reproducibly without causing artefacts, preserving image quality and comparability.

What you need

  • Arm, knee and head supports (foam), straps
  • Chest band (for the breast in MPI, per institution)
  • Patient gown, container for metal items
  • Gamma camera or PET/CT

Steps

  1. General: explain the procedure and its duration; remove metal objects; have the patient void just before imaging; make them comfortable and ask them to keep still.
  2. Bone: the patient drinks plenty and voids often; delayed images 2-5 hours after injection. Supine, body midline centred on the detector; bring the collimator as close to the patient as possible.
  3. Bone whole body: anterior and posterior images, 10-15 cm/min for delayed imaging; choose a speed giving more than 1.5 million counts per image. If the bladder overlies the pubis or sacrum, take an extra view after voiding.
  4. MPI SPECT: supine, arms (especially the left arm) above the head and supported; with a 180° orbit the left arm must be outside the field of view. Female patients remove the bra; breast position is kept the same for stress and rest.
  5. Prone imaging in MPI is optional: it reduces motion and inferior-wall attenuation; stress and rest are compared in the same position.
  6. PET/CT: supine, arms above the head and supported (arms down for head-and-neck imaging); this avoids CT truncation and beam-hardening artefacts.
  7. PET/CT for radiotherapy planning is acquired in the treatment position with the same immobilisation devices.
  8. Record the position, arm position and anything non-standard; use the same position for repeat studies.

Key points

  • Excess collimator-to-patient distance, motion, bladder activity, urine contamination and metal objects are frequent sources of error in bone scintigraphy.
  • If the patient cannot raise both arms for PET/CT, one arm can be up and the other by the side, or both close to the body; CT truncation is avoided.
  • Prone MPI does not eliminate attenuation but moves it; it can cause anteroseptal artefactual defects from sternal attenuation.
  • Position must be the same for stress and rest acquisitions and for serial PET scans; otherwise comparison is compromised.
  • Comfort reduces motion: knee and arm supports and a short explanation prevent most movement.

Common mistakes

  • Not removing metal objects
  • Not having the patient void before imaging
  • Leaving the left arm in the field of view in MPI
  • Different positions for stress and rest
  • Keeping the collimator far from the patient

Safety. Keep time close to the patient short; plan positioning before injection; use pads against urine contamination.

Sources: EANM kemik sintigrafisi kılavuzu (Eur J Nucl Med Mol Imaging 2016; doi 10.1007/s00259-016-3415-4, PMID 27262701) · EANM MPI SPECT ve SPECT/BT işlem kılavuzu (2015) · EANM Technologist's Guide: Myocardial perfusion imaging (2014) · EANM FDG PET/BT tümör kılavuzu 2.0 (2015) ve 3.0 (Boellaard ve ark., EANM J 2025;1:100006; doi 10.1016/j.eanmj.2025.100006). Last reviewed: 23 November 2026.