血漿採取法によるGFR
クリアランスは投与量 / 曲線下面積である。slope-intercept法は早期の分布成分を見落としてGFRを過大評価するため、Brøchner-Mortensenの式で補正する。
タイトルと要約は機械支援による翻訳(下書き)です。アニメーション本体は英語です。 英語の完全版ページ →
A GFR agent is injected into one arm and tubes are taken from the other arm at 120, 180 and 240 minutes; the area under the plasma concentration curve fills and clearance is computed as dose / area. In the second step the slope and intercept are drawn from the late samples on a log scale, the early area left out is shown, and the Brøchner-Mortensen correction is applied.
Key idea: clearance is dose / area under the curve; the slope-intercept method misses the early distribution component, overestimates GFR and is corrected with Brøchner-Mortensen.
Source: book (16.4: clearance = dose / area, slope-intercept with 2-4 samples, sampling from the other arm, Brøchner-Mortensen correction, Cr-51 EDTA, Tc-99m DTPA, when not to sample). Correction formula Brøchner-Mortensen 1972 (adults): Cl = 0.990778·Cl₁ − 0.001218·Cl₁². Dose and curve values are representative.
How to use: a single click on the drawing pauses and resumes; drag the time bar below to any moment; ← → change steps.