PETの2D収集と3D収集

セプタを外すとリング間の同時計数も数えられ、感度は約4–8倍になる。その代わりに散乱同時計数と偶発同時計数が増え、再構成も複雑になる。

タイトルと要約は機械支援による翻訳(下書き)です。アニメーション本体は英語です。 英語の完全版ページ →

An axial section shows two rows of detector rings with the patient between; annihilation photons leave at random axial angles. In 2D the septa between rings stop oblique and scattered photons; in 3D coincidences between all rings are accepted.

Continuous stream · no seeking

Key point: removing the septa lets cross-ring coincidences count and raises sensitivity ~4–8×; the price is more scatter and randoms and a more complex reconstruction.

Source: book (2D: same/adjacent ring, scatter ~15–20%; 3D: sensitivity ~4–8×, scatter ~30–40%, randoms↑). The drawing is schematic: the axial direction is stretched, so the ratio of counts in the drawing does not show the real gain. Photon directions isotropic; single scatter of ~30% of photons in the body is illustrative.

How to use: a single click on the drawing pauses and resumes; ← → switch between 2D and 3D.