PET二维(2D)与三维(3D)采集
去掉隔栅后,跨环(cross-ring)符合也被计数,灵敏度提高约4–8倍;代价是散射符合和随机符合增多,重建也更复杂。
标题和摘要为机器辅助翻译(草稿);动画内容为英文。 英文完整页面 →
An axial section shows two rows of detector rings with the patient between; annihilation photons leave at random axial angles. In 2D the septa between rings stop oblique and scattered photons; in 3D coincidences between all rings are accepted.
Key point: removing the septa lets cross-ring coincidences count and raises sensitivity ~4–8×; the price is more scatter and randoms and a more complex reconstruction.
Source: book (2D: same/adjacent ring, scatter ~15–20%; 3D: sensitivity ~4–8×, scatter ~30–40%, randoms↑). The drawing is schematic: the axial direction is stretched, so the ratio of counts in the drawing does not show the real gain. Photon directions isotropic; single scatter of ~30% of photons in the body is illustrative.
How to use: a single click on the drawing pauses and resumes; ← → switch between 2D and 3D.