PET 2차원 및 3차원 수집

격벽(septa)을 없애면 링 사이의 동시계수(coincidence)도 세어져 예민도가 약 4–8배 높아진다. 대신 산란과 무작위 동시계수가 늘고 재구성이 더 복잡해진다.

제목과 요약은 기계 보조 번역(초안)입니다. 애니메이션은 영어로 되어 있습니다. 영어 전체 페이지 →

An axial section shows two rows of detector rings with the patient between; annihilation photons leave at random axial angles. In 2D the septa between rings stop oblique and scattered photons; in 3D coincidences between all rings are accepted.

Continuous stream · no seeking

Key point: removing the septa lets cross-ring coincidences count and raises sensitivity ~4–8×; the price is more scatter and randoms and a more complex reconstruction.

Source: book (2D: same/adjacent ring, scatter ~15–20%; 3D: sensitivity ~4–8×, scatter ~30–40%, randoms↑). The drawing is schematic: the axial direction is stretched, so the ratio of counts in the drawing does not show the real gain. Photon directions isotropic; single scatter of ~30% of photons in the body is illustrative.

How to use: a single click on the drawing pauses and resumes; ← → switch between 2D and 3D.