콜리메이터의 종류와 거리의 영향

콜리메이터는 영상의 기하를 정한다. 평행구멍 콜리메이터(parallel-hole collimator)에서는 거리가 멀어질수록 공간 해상도가 나빠지지만 예민도는 그대로이다. 핀홀 콜리메이터(pinhole collimator)는 상을 뒤집고 가까운 물체를 확대한다.

제목과 요약은 기계 보조 번역(초안)입니다. 애니메이션은 영어로 되어 있습니다. 영어 전체 페이지 →

Photons from an object are selected by the collimator and those passing form the image in the crystal; parallel-hole, pinhole, converging and diverging collimators and the effect of source–collimator distance are shown.

Step 1/5

Role of the collimator

The collimator passes to the crystal only photons travelling along the holes and absorbs the others in the septa, so each point lands in its own place on the crystal and an image forms. The price is sensitivity: only a very small fraction of photons get through. The collimator is the main determinant of both resolution and sensitivity, which are inversely related.

Parallel-holeNo magnification · not inverted · distance↑ → resolution↓, sensitivity constantCrystalObjectImage
Upper half of the object orange, lower half blue. Photon numbers and angles are not to scale: in reality the collimator passes only a very small fraction of photons.
Continuous stream · no seeking

Key point: the collimator sets the image geometry; with a parallel-hole collimator resolution worsens with distance while sensitivity stays the same; a pinhole inverts the image and magnifies nearby objects.

Source: the collimator geometric-resolution relation; LEHR d = 1.5 mm, L = 35 mm (manufacturer value), Ri = 3.5 mm; Rc = d·(L + z)/L, Rs = √(Ri² + Rc²); pinhole M = f/z, f = 20 cm example value. Photon numbers and angles not to scale. Septa step: LEHR 1.5 / 0.2 / 35 mm and MEGP 3.0 / 1.05 / 58 mm (GE Discovery; Wright et al., Phys Med Biol 2017), HE 4.0 / 2.0 / 59.7 mm (Siemens Symbia); lead attenuation coefficients from NIST XCOM. This is a 2D cross-section simulation; true (3D) penetration fractions differ (Monte Carlo: LEHR + I-131 ~50%, HEGP + I-131 ~18%; Sawant et al., J Nucl Med Technol 2025).

How to use: a single click on the drawing pauses and resumes; in the distance steps drag the object or use ← → to move it closer or farther; in the septa step ↑ ↓ change the radionuclide.