准直器类型与距离的影响
准直器决定成像的几何关系;使用平行孔型准直器时,空间分辨率随距离增加而变差,而灵敏度保持不变;针孔型准直器使图像倒置,并放大靠近准直器的物体。
标题和摘要为机器辅助翻译(草稿);动画内容为英文。 英文完整页面 →
Photons from an object are selected by the collimator and those passing form the image in the crystal; parallel-hole, pinhole, converging and diverging collimators and the effect of source–collimator distance are shown.
Step 1/5
Role of the collimator
The collimator passes to the crystal only photons travelling along the holes and absorbs the others in the septa, so each point lands in its own place on the crystal and an image forms. The price is sensitivity: only a very small fraction of photons get through. The collimator is the main determinant of both resolution and sensitivity, which are inversely related.
Key point: the collimator sets the image geometry; with a parallel-hole collimator resolution worsens with distance while sensitivity stays the same; a pinhole inverts the image and magnifies nearby objects.
Source: the collimator geometric-resolution relation; LEHR d = 1.5 mm, L = 35 mm (manufacturer value), Ri = 3.5 mm; Rc = d·(L + z)/L, Rs = √(Ri² + Rc²); pinhole M = f/z, f = 20 cm example value. Photon numbers and angles not to scale. Septa step: LEHR 1.5 / 0.2 / 35 mm and MEGP 3.0 / 1.05 / 58 mm (GE Discovery; Wright et al., Phys Med Biol 2017), HE 4.0 / 2.0 / 59.7 mm (Siemens Symbia); lead attenuation coefficients from NIST XCOM. This is a 2D cross-section simulation; true (3D) penetration fractions differ (Monte Carlo: LEHR + I-131 ~50%, HEGP + I-131 ~18%; Sawant et al., J Nucl Med Technol 2025).
How to use: a single click on the drawing pauses and resumes; in the distance steps drag the object or use ← → to move it closer or farther; in the septa step ↑ ↓ change the radionuclide.