전신 PET(total-body PET)
전신 PET의 약 40배 예민도는 온몸을 한 번에 보는 것과 비스듬한 광자 쌍을 잡는 것 모두에서 오며, 선량 감소, 시간 단축 또는 화질 향상에 쓸 수 있다.
제목과 요약은 기계 보조 번역(초안)입니다. 애니메이션은 영어로 되어 있습니다. 영어 전체 페이지 →
Left: annihilations in a patient seen from the side and photon pairs caught or lost depending on the axial FOV, then conventional and total-body PET images building up. Right: total-body and single-organ geometric efficiency of four systems relative to conventional, and how sensitivity, activity, time and decay set the counts.
The ~40-fold sensitivity of total-body PET comes both from seeing the whole body at once and from catching oblique photon pairs, and it can be spent on lower dose, shorter time or better quality.
Source: FOV lengths, ~40-fold gain and ways of use from the book (28.4). Geometric efficiency computed on this page by Monte Carlo (no attenuation, scatter or acceptance-angle limit); phantom and counts representative; F-18 T½ 109.8 min.