全身PET(total-body PET)

全身PET约40倍的灵敏度既来自同时覆盖全身,也来自捕获斜向光子对;这一增益可用于降低剂量、缩短时间或提高图像质量。

标题和摘要为机器辅助翻译(草稿);动画内容为英文。 英文完整页面 →

Left: annihilations in a patient seen from the side and photon pairs caught or lost depending on the axial FOV, then conventional and total-body PET images building up. Right: total-body and single-organ geometric efficiency of four systems relative to conventional, and how sensitivity, activity, time and decay set the counts.

Continuous stream · no seeking

The ~40-fold sensitivity of total-body PET comes both from seeing the whole body at once and from catching oblique photon pairs, and it can be spent on lower dose, shorter time or better quality.

Source: FOV lengths, ~40-fold gain and ways of use from the book (28.4). Geometric efficiency computed on this page by Monte Carlo (no attenuation, scatter or acceptance-angle limit); phantom and counts representative; F-18 T½ 109.8 min.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the system's axial field of view in the first step and the way the sensitivity is used in the second; ← → change steps.